Note:
The installed plug-in cards can be displayed in the menu [Operating] - [System
information] - [Show HW options].
Show hardware options
@admin
Built-in
RS-232
Option-1
Option-2
Option-FB
WP A
PR5500/04
RS-485/RS-485
PR5500/17
Digital I/O
PR5500/W1
weighing electronic
-empty-
Info
Monitor
Showing hardware options
Description
Internal
RS-232
Interfaces (ixed installation)
Option 1
Option 2
PR 5500/04
PR 5500/19
RS-485/RS-485
Digital I/O
Slots for optional cards 1 and 2
FB option
-empty-
Slot for ieldbus cards,
in this case: [-empty-]
because a ieldbus card has not been
installed.
WP A
PR 5500/W1
Weighing elec-
tronics
Slot for WP A,
in this case: Internal weighing elec-
tronics
5.2.3.3
Terminal coding
The relevant plug connections will be coded when delivered according to the
combinations speciied when placing the order. The plug connections will only need to be
coded if plug-in cards are supplied subsequently.
The coding position for each of the plug-in cards is outlined in Chapter
.
Procedure
1
2
3
4
5
6
1. Remove the terminal (6) from the terminal strip (4).
2. Remove the coding pin (1) from the coding proile (2).
3. Insert the coding pin (1) into the corresponding slot (3) in the terminal strip (4).
5 Maintenance/repairs/soldering work/cleaning
Process Controller Maxxis 4 PR 5500
Minebea Intec
EN-101