![Minebea Intec EF-LT-3-P1 Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/minebea-intec/ef-lt-3-p1/ef-lt-3-p1_operating-instructions-manual_1801202056.webp)
This procedure is interruptable by the
(Exit) Button
The weight values are fix
— WP 2
Only visible if there is a second scale defined
in menu [OP.FUNC] - [WP2]
:
Initiates a linearization procedure for WP2
This procedure is interruptable by the
(Exit) Button
The weight values are fix
— GEO
Geographical Adjustment Factor (GEO) is used
to adjust the calibration based on the current
location
— 12
selectable area from 0 to 31 in single steps*
— END
Leave menu level
— D
Step with of the scale (1 digit) changeable via
or
Button only!
— CAL.WGT
Calibration weight, which is used in [CAL.ADJ] -
[CAL] - [WP2] and which is free selectable
within 30% Max-Load and Max-Load
— MAX.CAP
Maximum capacity of the scale
"H" is displayed if this entered value is reached.
— END
Leave menu level
— SER.NUM
show serial number (if two scales selected, then
show two serial numbers)
— W1.471
With
(No) or
Button next six
digits or first six digits again are displayed.
— W2.471
With
(No) or
Button next six
digits or first six digits again are displayed.
— END
— TYPE
Show scale type name
— SFT.VER
Show software version
— SR 2.64
— GEO.DAT
Show selected GEO area which is valid for both
scales
— 12
selected GEO area
— WP1
Info Wp1
— D
Accuracy
— MAX
Max.-Load
— END
— WP2
Info Wp2 when selected in menu
— D
Accuracy
7.4.9
Menu Selection [AD.CON2]
Enter this menu to ADC configuration for scale 2
AD.CON2
7.4.10
Menu Selection [INFO]
INFO
Compact Scale Puro® Count
7 Menu Settings
EN-54
Minebea Intec