11 - 11
■
Patient having frequent arrhythmia
■
Weak patient pulse
NOTE
•
The monitor always performs NIBP measurement at deflation stage if Patient
Category is Neo.
11.8.8
Setting Measurement Times
To set the measurement times, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
3.
Select
Parameters Setup
.
4.
Under
NIBP
tab, set
Measurements
to
Twice
.
11.8.9
Correcting the NIBP Measurements
The middle of the cuff should be at the level of right atrium. If the limb is not at the heart
level, you need to correct the measurement:
■
Add 0.75 mmHg (0.10 kPa) to the displayed value for each centimetre higher.
■
Deduct 0.75 mmHg (0.10 kPa) to the displayed value for each centimeter lower.
11.9
BP Averaging
In BP averaging mode, a group of BP measurement is performed on a patient
automatically. At the completion of measurement, a group of averaged BP readings is
displayed.
11.9.1
Enabling BP Averaging
BP averaging is available only when the
Department
is set to
Physician Office
and the
monitor is working in
Spot Check
mode.
To change the department, see
.
To enable BP averaging, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
3.
Select
Parameters Setup
→
the
NIBP
tab.
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...