46 MW306 Portable Meter
12. TROUBLESHOOTING
Symptom
Problem
Solution
Slow response /
Excessive drift
Dirty probe
Remove and clean the
sleeve. Make sure the
probe rings are clean.
Reading fluctuates
up and down (noise)
Probe sleeve placed
incorrectly.
Air bubbles inside
sleeve.
Correctly place the
sleeve.
Tap the probe to remove
air bubbles.
Display shows EC,
TDS or NaCl reading
blinking
Reading
is out of range
Recalibrate the meter.
Sample not within
measurable range.
Disable autoranging
feature.
Meter fails to
calibrate or gives
faulty readings
Broken probe
Replace the probe.
LCD tags displayed
continuously at
startup
ON/OFF key is
blocked
Check the keyboard. If
error persists, contact
Milwaukee Technical
Service.
“Internal Er X”
message
Internal error
Restart the meter.
If error persists, contact
Milwaukee Technical
Service.