32 MW306 Portable Meter
where:
R
T
ratio of sample conductivity to standard
conductivity at Temperature = (T) °C
C
T
(sample)
uncompensated conductivity at T °C
C (35, 15) = 42.914 mS/cm
the corresponding conductivity of
KCI solution containing a mass of
32.4356 g KCl/1 Kg solution
r
T
temperature compensation polynomial
a
0
= 0.008
a
1
= -0.1692
a
2
= 25.3851
a
3
= 14.0941
a
4
= -7.0261
a
5
= 2.7081
b
0
= 0.0005
b
1
= -0.0056
b
2
= -0.0066
b
3
= -0.0375
b
4
= 0.0636
b
5
= -0.01442
c
0
= 0.008
c
1
= 0.0005
X = 400R
T
Y = 100R
T
%NaCl Percentage
In this scale 100% salinity is equivalent to roughly 10% solids.
If the reading is outside the range, the full-scale value (400.0%)
will be displayed blinking.
Natural Sea Water
The Natural Sea Water scale extends from 0.00 to 80.00 g/L.
It determines salinity based upon a conductivity ratio of sample to
“standard seawater” at 15 °C.
where:
R
15
is the conductivity ratio
C
T
(sample) is uncompensated conductivity at T °C
C (35, 15) = 42.914 mS/cm is the corresponding conductivity of
KCI solution containing a mass of 32.4356 g KCl/1 Kg solution
r
T
is temperature compensation polynomial
Salinity (S) is defined by the following equation:
S = – 0.08996 + 28.2929729R
15
+ 12.80832R
15
2
–
– 10.67869R
15
3
+ 5.98624R
15
4
– 1.32311R
15
5
Note: The formula can be applied for temperatures between
10 and 31 °C.
Содержание MW306 MAX