SALINITY 33
9.4. WARNINGS & MESSAGES
Messages Displayed During Calibration
y
If an EC calibration is performed, the %NaCl calibration is
automatically cleared. A new %NaCl calibration is required.
y
If the reading exceeds expected calibration standard,
“WRONG STANDARD” message is displayed and calibration
can not be confirmed. Check if the correct calibration
solution has been used and / or clean the probe. See PROBE
MAINTENANCE section for details.
y
If the temperature is outside the 0.0 to 60.0 °C range, “WRONG
STANDARD TEMPERATURE” message is displayed. Temperature
value is displayed blinking.
Messages Displayed During Measurement
y
If the salinity measurement exceeds the specified limits or the
temperature exceeds (-20 to 120°C), “OUT OF SPEC” message
is displayed.