24 MW306 Portable Meter
Zero Calibration
For zero calibration, to correct readings around 0.00 µS/cm,
keep the dry probe in the air. The slope is evaluated when the
calibration is performed in any other point.
One-Point Calibration
1. Place the probe in the calibration solution making sure that
the sleeve holes are completely submersed. Center the probe
away from the bottom or beaker walls.
2. Lift and lower the probe to refill the center cavity and tap the
probe repeatedly to remove any air bubbles that may have
been trapped inside the sleeve.
3. Press CAL/EDIT to enter calibration. Use the
keys to select
a different standard value. The hourglass symbol and "WAIT"
message (blinking) are displayed until the reading is stable.
4. When reading is stable and close to the selected calibration
standard, SOL STD and ACCEPT tags are displayed blinking.