13. TROUBLESHOOTING
Symptoms
Problem
Solution
Reading fluctuates
up and down (noise)
DO probe electrolyte
contains air bubbles
Remove cap.
Refill, tap and reinstall.
Blinking DO reading
Reading is out of range
Remove cap.
Inspect and clean or replace if necessary.
Stir or increase flow rate.
Meter fails to calibrate
or gives faulty readings Broken probe
Replace the probe.
LCD tags displayed
continuously at startup ON/OFF key is blocked
Check the keyboard. If error persists,
contact Milwaukee Technical Service.
“Internal Er X”
Internal hardware error Restart the meter. If error persists, contact
Milwaukee Technical Service.