LOGGING
27
11. LOGGING
MW190
supports three types of logging: manual log on demand, log on stability and interval
logging. See Log Type in SETUP OPTIONS section.
The meter can hold up to 1000 log records. Up to 200 for manual log on demand, up to 200
for log on stability and up to 1000 for interval logging. See DATA MANAGEMENT section.
Note:
An interval logging lot can hold up to 600 records. When an interval logging
session exceeds 600 records, another lot file is automatically generated.
11.1. TYPES OF LOGGING
Manual log on demand
y
Readings are logged each time LOG/CLEAR is pressed
y
All manual readings are stored in a single lot (i.e. records made on different days
share the same lot)
Log on stability
y
Readings are logged each time LOG/CLEAR is pressed and stability criteria is reached
y
Stability criteria can be set to fast, medium or accurate
y
All stability readings are stored in a single lot (i.e. records made on different days are
logged in the same lot)
Interval logging
y
Readings are logged continuously at a set time interval (e.g. every 5 or 10 minutes).
y
Records are added to it until the session stops.
y
For each interval logging session, a new lot is created.
Note:
At the end of the logging session the meter returns to measurement screen.
A complete set of GLP information is stored with each log. See GLP section for details.