MW190 Dissolved Oxygen & Temperature Bench Meter
12. GLP
Good Laboratory Practice (GLP) allows the user to store and recall calibration data.
Correlating readings with specific calibrations ensures uniformity and consistency.
Calibration data is stored automatically after a successful calibration. GLP information is
included with every data log.
y
Press GLP/ACCEPT.
y
Use
keys to scroll through the calibration data displayed on the third LCD line.
y
Press ESC or GLP/ACCEPT to return Measurement mode.
12.1. DO INFORMATION
Calibration data displayed on the third LCD line:
y
Calibration standards and temperature
y
User selected altitude and salinity compensation values
y
Time, date
y
Calibration expiration time
“EXP WARN DIS” is displayed if calibration expiration time is disabled.
“NO CAL” is displayed blinking if no calibration has been performed (or calibration has
been deleted).