12. GLP
Good Laboratory Practice (GLP) allows the user to store and recall
calibration data. Correlating readings with specific calibrations
ensures uniformity and consistency.
GLP information is included with every data log. pH calibration
data is stored automatically after a successful calibration.
To view the pH calibration data:
y
Press GLP/ACCEPT when in Measurement mode.
y
Use the
keys to scroll through the calibration data
displayed on the third LCD line: Offset, slope, pH calibration
solutions, Time, date, calibration expiration time.
y
Press ESC or GLP/ACCEPT to return Measurement mode.
If calibration expiration time is disabled, “EXP WARN DIS” is
displayed.
If the instrument has not been calibrated or calibration has been
deleted, the blinking “NO CAL” message is displayed in GLP.