pH
27
For best results is recommended to:
y
Calibrate the probe before use and recalibrate periodically
y
Keep the electrode hydrated
y
Rinse the probe with the sample before use
y
Soak in
MA9015
Storage solution for at least 1 hour before
measurement
MTC Mode
When the probe is not connected the “NO T. PROBE” message is
displayed. The MTC tag and the default temperature (25 °C) with
blinking temperature unit are displayed.
1. Press CAL/EDIT and use the
keys to set the temperature
value manually.
2. Press GLP/ACCEPT to confirm or press ESC (or CAL/EDIT
again) to exit without saving.
Note: The temperature value used for MTC can be set only
when “NO T. PROBE” message is displayed.
8.4. WARNINGS & MESSAGES
Messages displayed during calibration
y
“WRONG BUFFER“ message is displayed blinking when the
difference between the pH reading and selected buffer value
is significant. Check if correct calibration buffer has been used.