8
MI605 Dissolved Oxygen Meter
7. ALTITUDE COMPENSATION
To enter altitude compensation, press the FACTOR key and the
meter will display the “ALT FACTOR” message. Use UP and DOWN
keys to set the altitude from 0 to 4.0 km, in steps of 0.1 km.
The meter will automatically compensate for altitude (see also the
table below for verification).
ºC
Altitude, Meters above Sea Level
ºF
0
m
300
m
600
m
900
m
1200
m
1500
m
1800
m
0
14.6
14.1
13.6
13.2
12.7
12.3
11.8
32.0
2
13.8
13.3
12.9
12.4
12.0
11.6
11.2
35.6
4
13.1
12.7
12.2
11.9
11.4
11.0
10.6
39.2
6
12.4
12.0
11.6
11.2
10.8
10.4
10.1
42.8
8
11.8
11.4
11.0
10.6
10.3
9.9
9.6
46.4
10
11.3
10.9
10.5
10.2
9.8
9.5
9.2
50.0
12
10.8
10.4
10.1
9.7
9.4
9.1
8.8
53.6
14
10.3
9.9
9.6
9.3
9.0
8.7
8.3
57.2
16
9.9
9.7
9.2
8.9
8.6
8.3
8.0
60.8
18
9.5
9.2
8.7
8.6
8.3
8.0
7.7
64.4
20
9.1
8.8
8.5
8.2
7.9
7.7
7.4
68.0
22
8.7
8.4
8.1
7.8
7.7
7.3
7.1
71.6
24
8.4
8.1
7.8
7.5
7.3
7.1
6.8
75.2
26
8.1
7.8
7.5
7.3
7.0
6.8
6.6
78.8
28
7.8
7.5
7.3
7.0
6.8
6.6
6.3
82.4
30
7.5
7.2
7.0
6.8
6.5
6.3
6.1
86.0
32
7.3
7.1
6.8
6.6
6.4
6.1
5.9
89.6
34
7.1
6.9
6.6
6.4
6.2
6.0
5.8
93.2
36
6.8
6.6
6.3
6.1
5.9
5.7
5.5
96.8
38
6.6
6.4
6.2
5.9
5.7
5.6
5.4
100.4
40
6.4
6.2
6.0
5.8
5.6
5.4
5.2
104.4