background image

Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the 

necessary information for the correct use of the 

instrument. Please read it carefully before using the 

meter. 

SPECIFICATIONS

Range

0.00 to 9.99 mg/L (NH

- N)

Resolution

0.01 mg/L

Precision

±0.30 mg/L @5.00 mg/L

Light Source

Blue LED 466 nm

Light Detector

Silicon Photocell and 466 nm narrow 

band interference filter

Method

Adaptation of Nessler method

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill the cuvet with 10 mL of sample, 

up to the mark, and replace the cap.

4.  Place the cuvet into the holder 

and ensure that the notch on the 

cap is positioned securely into the 

groove.

5.  Press ZERO and “SIP” will blink on 

the display.

6.  After a few seconds the display 

will show “-0.0-”. The meter is now 

zeroed and ready for measurement.

7.  Remove the cuvet. Open it.

 

 

10 ml

8.  Add 4 drops of NH

3

-1 (10 drops in case of seawater 

analysis) and swirl gently.

9.  Add 4 drops of NH

3

-2 (10 drops in case of seawater 

analysis). Replace the cap and swirl gently.

 

10 mL

 ▶

NH3 1

NH3 2

x 4

x 4

+

10. Immediately reinsert the cuvet into the holder and 

ensure that the notch on the cap is positioned 

securely into the groove.

Press and hold READ for 3 seconds and the display 

will show the countdown prior to the measurement. 

Alternatively, wait for 3 minutes and 30 seconds and 

just press READ. In both cases “SIP” will blink during 

measurement.

3’30”

11.  The instrument directly displays the concentration 

in mg/L of ammonia nitrogen (NH

3

-N).

12. To convert the reading to mg/L of ammonia (NH

3

), 

multiply by a factor of 1.216.

Interferences

 

y

Organic compounds like: chloramines, various 

aliphatic and aromatic amines, glycine, (positive 

error). To eliminate these interferences distillation 

is required.

 

y

Organic compounds like: aldehydes, alcohols (e.g. 

ethanol) or acetone (negative error). To eliminate 

these interferences distillation is required.

 

y

Sulfide (S²-): may cause turbidity.

 

y

Hardness above 1 g/L as Calcium Carbonate 

(CaCO

3

).

GUIDE TO DISPLAY CODES

This prompt appears for 1 second 

each time the instrument is turned 

on.
The dashes “- - -” indicates that 

the meter is in a ready state and 

zeroing can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.
“-0.0-”, the meter is in a zeroed 

state and measurement can be 

performed.
The blinking “BAT” indicates that the 

battery voltage is getting low and 

the battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication is 

displayed, the meter will lock up. 

Change the battery and restart the 

meter.
“Conf”, the meter has lost its 

configuration. Contact your dealer 

or the nearest Milwaukee Customer 

Service Center.

ERROR MESSAGES

On Zero Reading

 

y

Blinking “-0.0-” indicates that the 

zeroing procedure failed due to a 

low signal-to-noise ratio. In this 

case press ZERO again.

 

y

“no L”, the instrument can not 

adjust the light level. Please 

check that the sample does not 

contain any debris.

 

y

“L Lo”, there is not enough light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

 

y

“L Hi”, there is too much light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

Отзывы: