10• Wait 1 minute for free chlorine or 2,5 

minutes for total chlorine and press 
READ. In both cases “SIP” will blink 
during measurement. 

11• The instrument directly displays 

concentration in mg/L of free or total 
chlorine. 

 
Interferences: 
 

• Bromine (positive error) 
• Chlorine dioxide (positive error) 
• Iodine (positive error) 
• Oxidized Manganese (positive error) 
• Chromium (positive error) 
• Ozone (positive error)

 

For pH measurement 

12•   Fill a cuvet with 10 mL of unreacted 
 

sample up to the mark. 

13•  Add 5 drops of pH-0 

reagent. Replace  the cap 
and shake gently. 

14•  Insert the cuvet into the 

holder  and ensure that the 
notch of the 

cap is 

positioned securely into the 
groove. 

15•   Pres READ and “SIP” will  
 

blink during measurement. 

16•   The instrument dirrectly  
 

display the pH value. 

 
 

GUIDE TO DISPLAY CODES: 

This prompt appears for 1  
second each time the 
 instrument is turned on. 

Sampling In Progress. 
 Flashing “SIP” prompt  
appears each time 
the meter is performing a measurement. 

“-0.0-”, the meter is in 
 a zeroed state and  
measurement  can 
beperformed. 

The blinking “BAT” indicates  

thatb the battery voltage is  
getting low and the battery  
needs to be replaced. 

“-bA-”, the battery is dead 
and must be replaced. Once 
 this indication is displayed,  
the meter will lock up.  
Change the battery and restart the meter. 

“Conf”, the meter has lost  
its configuration. Contact 
 your  dealer or Milwaukee 
Instruments.

 

ERROR MESSAGES 

On zero reading 

•  Blinking “-0.0-” indicates  
  that the zeroing procedure 
  failed due to a low signal- 

tonoise ratio.  In this case 
press ZERO again. 

• “no L”, the instrument  

cannot adjust the light level.  Please check that 
the sample does not contain any debris. 

•  “L Lo”, there is not enough light  to perform a 

measurement.  Please check 
the preparation of the 
zerocuvet. 

•  “L Hi”, there is too much light 

to perform a measurement. 

  Please check the preparation 
  of the zero cuvet.

 

On sample reading 
 

•   “-SA-”, there is too much  
  light 

for the sample 

  measurement. 

Please 

  check if the right  sample 
  cuvet is inserted. 
•  “Inv”, the sample and 
  the zerocuvet are inverted. 
•  “ZErO”, a zero reading was 

not taken. Follow the 
instruction 

in the 

measurement procedure  
for zeroing themeter. 

•  Under range. A blinking “0.00” 

indicates that the sample 
absorbs less light than the 

  zero reference. Check the 
  procedure and make sure you 
  use the same cuvet for reference 
  (zero) and measurement. 
•  A flashing value of the  
  maximum concentration 
   indicates an over range  
  condition. The concentration of the  
  sample is beyond the  

programmed  range: dilute the sample and  re-
run the test.

 

 

BATTERY REPLACEMENT 

Battery replacement must only take place in a 
non-hazardous environment.  Simply rotate the 
battery cover on the back of 
the meter.  Detach 
the   battery from the 
terminals and attach  
a  fresh 9V  battery 
while 

paying 

attention to 
the correct polarity. 
Insert the battery 
and replace the 
cover. 

 

ACCESSORIES: 

MI526-100

 for free chlorine (100 tests) 

MI524-100

 for total chlorine (100 tests) 

MI509-100 

pH reagent (100 tests) 

MI511-100 

Free & Total Chlorine and pH reagent 

set (100 tests) 

MI0001 

Glass cuvets (2 pcs) 

MI0002 

Caps for cuvets (2 pcs) 

MI0003 

Stoppers for cuvets (2 pcs) 

MI0004 

Tissue for wipping cuvets (4 pcs) 

MI0005 

9V battery (1 pc) 

 
WARRANTY 
This instrument is warranted against defects in materials 
and manufacturing for a period of two years from the date 
of purchase. 
If during this period the repair or replacement of parts is 
required, where the damage is not due to negligence or 
erroneous operation by the user, please return the meter 
to either distributor or our office in the original packing and 
the repair will be free of charge. 
Damages due to accidents, misuse, tampering or lack of 
prescribed maintenance are not covered.

 