OPERATION: 

• Connect  the  supplied  12

VDC  power  adapter  to  the
meter and to the mains line.

• Connect the ORP electrode

to  the  BNC  socket  on  the
bottom of the meter.

• Always  remove  the  elec-

trode  protective  cap  before
taking  any  measurement.  If
the  electrode  has  been  left
dry, soak the tip (bottom 2.5
cm)

in 

rinse 

solution

(M10000B) for a few minutes to reactivate it.

• Turn the instrument on by pressing the ON/OFF key.
• Allow  the  reading  to  stabilize  and  the  meter  will  start  continuous

monitoring.

• The meter is factory calibrated.
• A blinking alarm will indicate when the measured ORP value is lower or

higher than the setpoint, depending on user selection.

NOTE: 

• The  power  contact  is  not  protected  by  fuse  inside  the  meter.  It  is

recommended to protect it outside, against failure.

• WARNING:  when  the  alarm  led  is  blinking,  the  mains  voltage  is

present in the output power socket.

SETPOINT: 

The  setpoint  can  be  selected  by
adjusting the central front knob to the 
desired value.

The  selectable  range  for 

MC510

  is

from 0 to 600 mV.

The nature of the setpoint can be selected by setting the switch to the 

desired position (ABOVE or BELOW).

INSTALLATION PROCEDURE: 