7.3
Capturing a multi-color z-stack
1
Select
Measurement Mode > Multi Color Acquisition
in the
Measurement Wizard
window.
2
Select
xyz-Table Z
and
UltraII Filter
in the
Devices
list.
3
Click the
Autosave
button at the end of the row under
AS
. It will turn green.
4
Select the correct excitation and emission wavelengths for your sample under
Filter for Measurement
in
the
Settings 1
window. The selected checkboxes are used.
5
Move the slider under
Laser Transmission Control
while the filter is highlighted to set the laser power for
each individual filter.
NOTICE!
Higher laser power will bleach the sample.
6
Select the shortest emission wavelength, and create a sharp picture.
7
Correct the different focus lengths for different emission wavelengths using the
Chromatic Correction
button, starting with the next longer wavelength.
8
If all emission wavelengths have been corrected, click
Save Settings
in the toolbar.
9
To start the measurement, proceed as described under
Capturing an image on the previous page
.
Содержание UltraMicroscope II
Страница 1: ...UltraMicroscope II Imaging System User manual...
Страница 14: ...1 IMPORTANT SAFETY INFORMATION 12...
Страница 20: ...2 INTRODUCTION 18...
Страница 30: ...5 PREPARING THE INSTRUMENT 28...
Страница 38: ...6 LASER ALIGNMENT 36...
Страница 46: ...7 SETTING UP AN EXPERIMENT 44...
Страница 48: ...8 MAINTENANCE 46...
Страница 54: ...10 TECHNICAL SUPPORT 52...
Страница 57: ......