3
Fill the cuvette up to the notch (2) on the side with the selected imaging solution.
4
Place the cuvette securely on its holders again by holding the piston.
NOTICE!
Be careful not to spill any imaging solution.
5
Lower the piston carefully until the bottom of the microscope chamber is reached.
6
Fasten the cover plate by tightening the screw.
7
Attach the objective lens.
5.2
Setting the parameters for the alignment
1
Start the ImSpector software.
2
Select the appropriate
Instrument Mode
in the
Measurement Wizard
window for the ordered
configuration, such as
Zoom Body
.
3
Select the excitation and emission wavelengths under
Filter for Measurement
(
Channel Setup
in
software version 7.4.0 or later) in the
Settings 1
window of the
UltraII
view. For the alignment tool, the
excitation is 561 nm. The selection must be highlighted to become active.
4
Set
Laser Transmission Control
. A value between 5% and 10% is recommended.
5
P
R
EP
A
R
IN
G
TH
E
IN
ST
R
U
M
EN
T
26
Содержание UltraMicroscope II
Страница 1: ...UltraMicroscope II Imaging System User manual...
Страница 14: ...1 IMPORTANT SAFETY INFORMATION 12...
Страница 20: ...2 INTRODUCTION 18...
Страница 30: ...5 PREPARING THE INSTRUMENT 28...
Страница 38: ...6 LASER ALIGNMENT 36...
Страница 46: ...7 SETTING UP AN EXPERIMENT 44...
Страница 48: ...8 MAINTENANCE 46...
Страница 54: ...10 TECHNICAL SUPPORT 52...
Страница 57: ......