General installation and maintenance instructions
HOSPITAIR PACK S (MV) - 11 bar
Manual N° 520003 version 5
12/2016
Page 6/17
3.4.2. Adsorption
dryer
The oxygen generator gives the order for the adsorption dryer to start. As long as the generator is active, the adsorption
dryer operates.
When the oxygen generator gives the order for the adsorption dryer to stop, the dryer will stop at the end of the cycle.
3.4.3.
Forced running mode for compressor/dryer
The compressor is equipped with a button to choose the automatic running mode or the forced running mode.
The automatic running mode is corresponding to the order given by the oxygen generator.
3.4.4.
Emergency stop compressor
Use the emergency button only in extreme necessity because it stops suddenly the compressor and if the
compressor is on charge, oil can go back in suction.
During test, it is better to use the I/O button of the SIGMA CONTROL to stop the compressor.
3.4.5. Transport
chocks
Don't forget to take off the transport chocks.
3.4.6.
Information about shot risk
The oil separator, mounting between gaskets, can be isolated from the ground of the unit.
It can be load electro statically and there is a risk of spontaneous ignition of the oil vapour. The oil separator gaskets
are equipped with aluminium braid to maintain the ground.
This situation is exceptional but you must be informed to take safety precaution to operate and to use genuine material.