General installation and maintenance instructions 

HOSPITAIR PACK S (MV) - 11 bar 

 

Page 11/17 

12/2016 

Manual N° 520003 version 5 

 

SECTION 6.  COMMISSIONING 

6.1. First 

step 

 

Check the oil level of compressors. 

 

Before starting the equipment, first read all the running operation and maintenance manuals (screw compressor, 
ADSORMIL'S air dryer, PRO

2

XY). 

6.2. Switching 

on 

Each screw compressor is equipped with a rotation direction controller indicates phases are in correct position. If not, 
the compressor can't start. 
 

 

Shut off the network outlet valve. 

 

Start the compressor pushing the "

I

" key on the SIGMA CONTROL: if there is an electrical problem and fault, the 

compressor stops immediately. In this case, insulate the compressor electrically and cross 2 wires at compressor 
motor level. 

 

Restart the compressor to allow the pressure to go up to maximum level. 

 

Check the links for leaks. 

 

Cut the compressor and vent the receiver to 2 bar. 

 

Restart the compressor and open the network outlet valve gradually. 

 
 
 

SECTION 7.  CLEANING AND MAINTENANCE 

IMPORTANT: 
For this operation, use all necessary protective measures (gloves, mask, overalls etc.) then 
destroy contaminated consumables (antibacterial cartridges, condensates etc.) in according with 
that country’s applicable laws (regulation incinerator). 

 

Maintenance is to be entrusted to trained and approved personnel by the exploiting, with all 
necessary knowledge and acquainted with best practice. 

 
 

 Not to carry out maintenance operations when the device is in use. 

 

If the spare parts used during maintenance operations are not corresponding to the 
manufacturers' requirements, the manufacturer declines all responsibility in the event of an 
incident. 

 

The modification of the central units is forbidden. 

 

Modification of the device is forbidden without the approval of the manufacturer. 

 

If the central is modified following manufacturer agreement, a control and a suitable test must be 
realized to make sure that the central is always safe to use. 

 
To avoid risk of damage or prevent equipment become dangerous is essential use the compatible components provide 
by the manufacturer. 
 
 
 

Содержание HOSPITAIR PACK S MVA 8

Страница 1: ...RULES see manual n 511927 Spare parts list and maintenance instructions for compressors First apposition of the EC marking 2009 This document must not be disclosed or reproduced without permission fr...

Страница 2: ...ESCRIPTION OF THE OPERATION 5 INSTALLATIONS 7 3 5 DIMENSIONS 7 SECTION 4 ELECTRIC POWER SUPPLY 8 4 1 COMPRESSOR POWER SUPPLY 8 4 2 CONNECTION BETWEEN COMPRESSOR AND PROCOM 2 OF PRO2XY 8 4 3 SUMMARY OF...

Страница 3: ...in a safe place and to dispose of it with due care for the environment Dustbin on wheels barred of a cross symbol This symbol indicates that the equipment should not be thrown with domestic waste and...

Страница 4: ...2 Specification plates Type Hospitair units Serial number Air dryer type Serial number SECTION 2 CHOICE OF ROOM The unit must be located in a clean non humid non saline dustless and ventilated local w...

Страница 5: ...38 37 31 820 1 00 13 500 700 x 700 The person in charge of the installation has overall responsibility of the product environment requirements He is also in charge of setting up methods of struggle ag...

Страница 6: ...he pressure before the air treatment unit Then the air enters the filtration chain see the description in the manual concerning the filtration chain At the filtration chain output the air is then stor...

Страница 7: ...order given by the oxygen generator 3 4 4 Emergency stop compressor Use the emergency button only in extreme necessity because it stops suddenly the compressor and if the compressor is on charge oil c...

Страница 8: ...s HOSPITAIR PACK S MVA 8 MVB 12 MVB 16 HOSPITAIR PACK S MVC 19 MVC 23 HOSPITAIR PACK S MVD 31 MVE 38 MVE 46 3 5 2 Vertical air receivers 500 1000 litres E B A C F 1 2 Soupape Outlet Inlet Pressure gau...

Страница 9: ...mpliance with the legislation and regulations in force in your country The wire sections are given for information to 5A mm2 Even so the sections shall be defined by taking contacts with the wire prov...

Страница 10: ...report The connection between Alarm report on automate units must be done by the person in charge of the Installation He must control the compliance wiring and proper operation during the final accept...

Страница 11: ...rainage system OWAMAT separator used to separate oil and water In compliance with Law No 76 663 dated 19 July 1976 relating to installations classified for protecting the environment Law 92 3 dated 3...

Страница 12: ...pen the network outlet valve gradually SECTION 7 CLEANING AND MAINTENANCE IMPORTANT For this operation use all necessary protective measures gloves mask overalls etc then destroy contaminated consumab...

Страница 13: ...to stop Check the compressor discharge time before it stops completely 10 to 15 seconds 7 3 Safety relief valve When a relief valve with discharge line operates and must be tested no persons may stand...

Страница 14: ...to avoid all forced and tensions and to damp out all vibrations even non visible The base of the receiver will have to support the weight of the receiver including in situation of tests In performing...

Страница 15: ...g the compressed air unit object of this technical manual are protected by relief valves as indicated in the table hereafter Component Pressure setting bar Location of the safety relief valve Compress...

Страница 16: ...should assure that it is used in such an electromagnetic environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2...

Страница 17: ...ufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey 1 should be less than the compliance level...

Страница 18: ...o the maximum output power of the communications equipment Separation distance according to frequency of transmitter Rated maximum output power of transmitter W m 150 kHz to 80 MHz d 1 16 P 80 MHz to...

Страница 19: ......

Страница 20: ......

Отзывы: