General installation and maintenance instructions
HOSPITAIR PACK S (MV) - 11 bar
Manual N° 520003 version 5
12/2016
Page 14/17
SECTION 9. TEST PRESSURE
Each equipment under pressure was tested individually by its respective manufacturer.
The connection of the equipment is carried out by assembly without welding.
SECTION 10.
PROTECTION AGAINST ONVERPRESSURE
The various modules constituting the compressed air unit, object of this technical manual, are protected by relief valves
as indicated in the table hereafter:
Component
Pressure setting (bar)
Location of the safety relief valve
Compressor
13 bar
In the compressor
Dryer
---
Protected by the safety relief valve of the compressor
Receiver
11 bar
On the receiver
SECTION 11.
CLEANING
11.1. Units
cleaning
Do not use corrosive product
If the units system requires cleaning, use a soap solution and a clean cloth.
Then wipe well with a dry cloth.
For disinfection, use swabs soaked in a disinfectant solution.
Frequency: as required.