49
51.
Connect cable 4 to the main electronic board.
This cable comes from the rotor motor reducer block, goes to the right of the stator motor
reducer block, then through the cable holder located at the right of the electronic board,
and is connected to the electronic board.
4
4
4
52.
Ensure that no cable passes near the belts of the two motor reducer blocks or under the
two motor reducer blocks.
53.
Position the pump bottom plate underneath the pump housing.