36
35.
Remove the 2 threaded stems.
36.
Use a #5 hex key to install the 3 other pump housing screws, with washers.
37.
Perform a tightness test of the isolator before proceeding.
If this test fails, ensure that all the 5 screws on the pump housing are properly installed,
and ensure that the flat seal is not damaged. If the flat seal is damaged, replace it by a
new one (see
Accessories and Replacement Parts
).
38.
Place the stator motor reducer block in its housing by aligning the coupling pieces located
on the stator motor reducer block and at the bottom of the pump upper part.