Millipore Sigma PHCC30000 Скачать руководство пользователя страница 9

00110230w  Rev 10/20 

9 of 24

TIP

Having Trouble?   

See 

Troubleshooting

Creating a Wi-Fi

®

 Configuration File   

 

A “Wi-Fi

®

 configuration file” is needed for set up or to change Wi-Fi

®

 networks. 

On a computer, open up the text editing program. The examples below are using Microsoft

®

 Windows

®

 Notepad.

1.  Type following parameters with no spaces, as 

shown below: Replace 

exampleNetwork

 with the 

name of the desired wireless network. Replace 

ssidPassword

 with the password to the desired 

wireless network:

Name=

exampleNetwork

 

Passphrase=

ssidPassword

 

Type=wifi 

Favorite=true 

AutoConnect=true

2.  Save and name the file the same name of the 

desired wireless network with “.config” file 

extension.

exampleNetwork

.config

Note: 

If you’re using Notepad included with 

Microsoft

®

 Windows

® 

software, it will try to save 

the file as a .txt file by default. For this file to  

be saved properly, ensure it is being saved as  

All Files type like the example (right).

3.  Drag the file from the computer onto the USB 

thumb drive. Once the file has loaded, remove  

the USB thumb drive from the computer port. 

 

 

Connect to Wi-Fi

®

 Network

1.  From the Main Menu, select 

Settings

 and press the OK button. Select 

Wi-Fi

®

 and press the OK button. 

Insert the USB thumb drive containing the credential file into the Scepter™ 3.0 USB port. 

2.  Scroll to 

Turn On

and press the OK 

button. Scroll to 

Show 

Networks

, press the OK 

button. The instrument 

will begin 

Searching for 

networks...

 . 

3.  The instrument will list 

networks within range. 

Scroll to the desired 

network and press the OK 

button. When connection 

is complete, the screen 

will confirm successful 

connection. 

4.  Press the OK button to 

return to Settings Menu.

5.  Remove USB thumb 

drive.

01-Feb-18 14:16

Wifi

Back

Turn On

<

Press OK to accept

            

01-Feb-18 14:16

Wifi

Back

Turn Off

Show Networks

<

Press OK to accept

  

 

 

01-Feb-18 14:16

Wifi

Searching . . .

Searching for 

networks . . .

    

01-Feb-18 14:16

Wifi Networks

Back

 Guest

 MP123

 Sizzafritz

<

Press OK to accept

            

01-Feb-18 14:16

Wifi

Connecting . . .

Connecting to 

network:

MP123

   

01-Feb-18 14:16

Wifi

Press OK to continue

Connection was 

successful to

network . . .

MP123

                                  

Содержание PHCC30000

Страница 1: ...perates as MilliporeSigma in the U S and Canada The life science business of Merck KGaA Darmstadt Germany operates as MilliporeSigma in the US and Canada SigmaAldrich com User Guide Scepter 3 0 Handhe...

Страница 2: ...r computer easily No more cables The new Charging Station stores Scepter 3 0 out of the way when not in use freeing up valuable hood or lab bench space This system is intended for research use only Co...

Страница 3: ...nstrument to vibrations Vibrations may cause instrument malfunction or damage Do not autoclave or expose to high temperature To avoid damage to internal electronic components do not spray instrument w...

Страница 4: ...s the sensor Tech Service Port For Service use only USB Port Connects instrument to USB thumb drive Ergonomic Hook Allows comfortable handling of the instrument Secures the instrument to charger Contr...

Страница 5: ...the Recharging screen when placed in the Charging Station the following steps must be followed to start charging the Scepter 3 0 battery 1 Place Scepter 3 0 into the Charging Station 2 While on the s...

Страница 6: ...kable surface 1 Use double sided mounting tape 2 Remove tape liner from one side of tape and affix to back of mounting plate as shown below 3 Remove remaining tape liner and press firmly onto glass me...

Страница 7: ...14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 01 Feb 18 14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 1 Scroll to Edit Time and press the OK button 2 Scroll to...

Страница 8: ...nd select Properties If the File system is identified as FAT32 no formatting is needed Continue to Creating a Wi Fi Configuration File NOTE This view will also show how much space is available on your...

Страница 9: ...file from the computer onto the USB thumb drive Once the file has loaded remove the USB thumb drive from the computer port Connect to Wi Fi Network 1 From the Main Menu select Settings and press the...

Страница 10: ...were last set by the user either during a count or during post count data management 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Settings Back Date Time Bluetooth Wifi Gating M...

Страница 11: ...epter 3 0 instrument Do not use sensor if visibly damaged Attach the Sensor 1 When the Scepter 3 0 is powered on it will show Attach the Sensor to begin 2 Holding the sensor with the sensor size numbe...

Страница 12: ...net is recommend use low vacuum 5 Hg Cell Bead and Particle Suspensions Materials Required Cell sample Scepter 3 0 Test Beads Cat No PHCC3BEADS or particle suspensions Note Some particle suspensions...

Страница 13: ...ion and diameter or volume will be displayed on the screen Press the OK button to save up to 999 histograms in the Scepter 3 0 memory and open the Histogram Explorer Press the OK button to return to H...

Страница 14: ...03 Dec 17 Count 0 900 36 24 30 18 12 6 Dia in m 3 342e5 15 685 mL m Conc Dia 01 Feb 18 14 16 Histogram Edit Gates High gate Done 26 575 Low gate 12 585 Count 0 900 36 24 30 18 12 6 Dia in m Switch to...

Страница 15: ...wirelessly Materials Required Wireless Network broadcasting at 2 4 Ghz PC Mac or UNIX Computer connected to the same Wi Fi Network as Scepter 3 0 Scepter 3 0 Instrument Follow Wi Fi Network p 8 direct...

Страница 16: ...MS Press the OK button from any histogram on screen to open the Histogram Explorer menu 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Histogram Test 3 03 Dec 17 Count 0 900 36 24...

Страница 17: ...o printer Help Screens Tutorials 1 From the Menu screen scroll to Help press the OK button 2 Scroll to the topic of interest press the OK button 01 Feb 18 14 16 Scepter Software 4 6 01 Feb 18 14 16 Me...

Страница 18: ...ration of cell sample is within recommended operating range See Cell Bead and Particle Suspensions Test sample using Scepter 3 0 Test Beads Cat No PHCC3BEADS to ensure unit operating properly for conc...

Страница 19: ...y needs to be re charged Place Scepter 3 0 on charging stand to recharge the instrument See Battery p 5 Battery is not charging Ensure that the unit is correctly positioned in charging stand An animat...

Страница 20: ...ressing of OK button Fully submerge sensor in sample before pressing OK button Air bubble in sample Mix sample gently to avoid air bubbles Sensor lifted from sample before screen displays Sample Loade...

Страница 21: ...ed on FAT32 formatted USB thumb drive and try again See USB Thumb Drive Formatting p 8 Software update blocked on battery low Insufficient battery to perform update Recharge Scepter 3 0 and try again...

Страница 22: ...ntal Conditions Temperature 15 C to 25 C Maximum Relative Humidity 20 80 non condensing Storage Conditions Temperature 15 C to 30 C Maximum Relative Humidity 20 80 non condensing Scepter 3 0 Sensor Ma...

Страница 23: ...p E GP Sterile Vacuum Filtration System Millipore Express PLUS PES 0 22 500 38 12 SEGPU0538 0 22 500 45 12 SEGPU0545 0 22 1000 38 12 SEGPU1138 0 22 1000 45 12 SEGPU1145 Steritop E GP Sterile Funnel le...

Страница 24: ...Merck KGaA Darmstadt Germany and or its affiliates All Rights Reserved Global Registration Marks of Conformity USA TFB 1003 EUROPE C TAIWAN JAPAN R209 J00212 SOUTH KOREA R C Emd PHCC30000 RUSSIA BELAR...

Отзывы: