Page 18
MV-R 100
Version 2.5en - 12/2020
Installation
1.
Make sure the installation surface is horizontal.
2.
Place the
MV-R 100
on the installation surface.
$
Ensure that the surface is non-slip.
$
Lift and carry the sintering oven only by the bottom of the device.
Electrical connection
Building installation
;
The
MV-R 100
requires its own electrical circuit.
;
The electrical circuit in the building must have a fuse device of
minimum 16 A type K, Z (other fuse types in accordance with
regulation in the country of use).
;
An additional circuit breaker (with a tripping current of 30 mA) must
also be installed.
;
For safe electrical operation, the
MV-R 100
requires a protective
conductor connected to the socket.
;
When selecting the installation location, it must be noted that the
mains cable supplied is 2.0 m long and cable extensions are not
permitted. The power supply must be in the nominal voltage range of
200 - 240 volts.
Danger
Electrical energy!
Risk of death from electric shock.
h
Do not touch live cables and components with wet hands.
h
Observe the accident prevention regulations for working with electric
current.
h
Connect the device only to a power supply that matches the
information on the rating plate.
h
Removable mains cables must not be replaced with mains cables of
inadequate size.
4