HYB-1000 KIT
The HYB-1000 uses six AA, 1.5-volt batteries (alkaline
recommended) to allow testing of batteries down to 1 volt
and supply power while the menu is active. The analyzer can
test batteries down to 5.5 volts when the internal batteries are
not functioning.
Battery Power Indicator
The square in the upper left corner of the display indicates the
charge level of the battery pack. The square is black when the
battery pack is fully charged. It gradually changes to white as
the charge level declines. The HYB-1000 will display a warning
message when the batteries need replacing.
Full
Powered
by test
battery
Decreasing
Low
Replacing the HYB-1000 Batteries
1. Turn the HYB-1000 face down.
2. Remove the screw.
3. Press gently on the ridges above the arrow on the battery
compartment cover.
4. Slide the cover in the direction of the arrow and remove
the cover.
5. Remove the discharged batteries.
6. Insert new batteries as shown in the figure (right). Make
sure the positive and negative terminals are positioned
correctly.
7. Insert the battery door tabs into the slots on the analyzer
and slide the door closed, making sure the latch locks.
8. Insert and tighten the screw.
Latch
Press
here
Slide in
this direction
Screw
Chapter 10: HYB-1000 Internal Batteries