50
⚫
Greater than the time value (>)
For example, for positive pulse, if it is set as T>80ns, the trigger will happen stably only when the pulse
width is greater than 80ns (Figure 5-16 Trigger Time T>80ns).
Figure 5-16 Trigger Time T>80ns
⚫
Equal to the time value (=)
For example, for positive pulse, if it is set as T=80ns, the trigger will happen stably only when the pulse
width is equal to 80ns (Figure 5-17 Trigger Time T=80ns).
Figure 5-17 Trigger Time T=80ns
⚫
Not equal to the time value (≠)
For example, for positive pulse, if it is set as T≠80ns, the trigger will happen stably only when the pulse
width is not equal to 80ns (Figure 5-18 Trigger Time T≠80ns).
Figure 5-18 Trigger Time T≠80ns
The trigger pulse width time can be set as 8ns~10s.
Tap the pulse width time setting box
to pop up the time adjustment interface (as shown in
Figure 5-19), and adjust the pulse width time. Adjust the pulse width time by adjusting or dragging the time
scale.