TIMING Tab
5. If an Input Reference becomes available again, the Hardware Clock synchronizes with it
and the time quality character becomes " " (time valid) again.
Usually there is a short delay between the Hardware Clock changing state and the time qual-
ity character changing.
Here are some potential guidelines for configuring the Flywheel Quality Character (FQC):
n
The user sets the FQC to " " if one or more of the following are true:
n
The SyncServer is configured with two or more synchronizing NTP associations
and the user is satisfied with using time from other NTP associations.
n
The SyncServer oscillator type has superior time keeping properties compared to
the receiving equipment. This is usually the case since most computer equipment
uses uncompensated quartz oscillators.
n
The user sets the FQC to "F" if the receiving equipment can handle "F" as a time quality
character in some way that is useful and distinct from the " " or "X" time quality char-
acters.
n
The user sets the FQC to "X" so that the receiving equipment to handles time from NTP
or the Hardware Clock internal oscillator as "X" (time invalid).
Troubleshooting: If the time quality character remains "X" (time invalid) even though Input
References are connected to the SyncServer.
n
Verify that the physical connection to the input connector is valid and that there are no
cable breaks or short circuits.
n
On the
TIMING - HW Clock
page, verify that the Input Reference is
Enabled
and that
Forced
Timing Source
is set to
Auto
.
n
For IRIG-B, on the
REFERENCES - IRIG-B
page, check that the
Timecode Input
setting
matches the input signal type.
n
For GPS, wait for the GPS receiver to complete the GPS acquisition process and
achieve "locked" status. Also see
Operating in "Window Mode"
(on page 102).
Also see
TIMING - Holdover
(on page 45) and TIMING - HW Clock.
Note: If the user sets
Forced Timing Source
on the
TIMING - HW Clock
page to
Free Run
, the
Flywheel Quality Character
in effect at that moment remains in effect thereafter.
997-01520-02 Rev. G2
..........................................................................
Page 49
Содержание SyncServer S200
Страница 1: ...SyncServer S200 S250 S250i User Guide Rev G2 May 2015 Part Number 997 01520 01 CD Number 998 01520 01...
Страница 2: ...Page 2...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 78: ......
Страница 114: ......
Страница 124: ......
Страница 178: ......
Страница 182: ...Glossary Page 182 997 01520 02 Rev G2...
Страница 190: ......