LVS-9580/9585 Operating Instructions - English
LVS-9580/9585 Operating Instructions - English
Page 11 of 27
3. Firmly grip the LVS-9580/9585 handle. There is no need to pull the trigger at this time (pulling the trigger
causes the LVS-9580/9585 to capture a live image; however, the image is automatically live while the
system is in Calibration mode).
4. Place the LVS-9580/9585 window over one of the Master Grade barcodes making sure the four rubber
feet surrounding the window rest firmly on a flat, stable surface. The rubber feet hold the test card in
place and prevent movement of the test card. See example of Master Grade barcodes above. It is
important to note that not all of the rubber feet will fit onto the calibration card.
Not all of the
rubber feet will fit
onto the
calibration card.
Verifier Window