Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
- guidance
Voltage dips
(IEC
61000-4-11)
0%
U
T
for 0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
for 1 cycle
and
70%
U
T
for 25 cy-
cles,
50Hz for 30 cy-
cles,
60 Hz
singlephase at 0°
Non
applicable
Mains power quality should be that of
a professional health care
establishment.
If operation of the system requires
continued use during power cuts, it is
recommended that the medical device
be powered by a separate power
source (UPS, etc.).
Voltage
Interruptions
(IEC
61000-4-11)
0 %
U
T
;
for 250 cycles
at 50 Hz
for 300 cycles
at 60 Hz
Non applica-
ble
Mains power quality should be that of
a professional health care
establishment.
If operation of the system requires
continued use during power cuts, it is
recommended that the medical device
be powered by a separate power
source (UPS, etc.).
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURER'S
DECLARATION: ELECTROMAGNETIC IMMUNITY, PORTABLE RF
EQUIPMENT
WARNING:
Portable RF communication devices (including peripherals such as antenna
cables and external antennae) should not be used closer than 30 cm (12 inches) to any part
of the Spider SAS, including the cables specified by the manufacturer. Otherwise, the
performance of these devices may be impaired.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environ-
ment - guidance
Electromagnetic
fields
Radiated RF
(IEC 61000-4-3)
10 V/m
80 MHz – 2.7
GHz
80 % MA at
1 kHz
10 V/m
80 MHz – 2.7
GHz
80 % MA at 1 kHz
Professional health care
establishment and home
health care environment
16.5.
16. APPENDIX 2: ELECTROMAGNETIC COMPATIBILITY
56
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F