PPS-
20
Piezo Positioner Stage
10
Rev: 3.12
MICRONIX USA,LLC
www.micronixusa.com
Reference Manual
For a PPS-20 with a digital encoder, connect the stage as shown below
using the supplied 15 to 9 pin cable to connect the Encoder Module to the
Controller.
5.2
18B
Vacuum Environments
5.2.1
32B
Handling and Preparation
When preparing the stage for vacuum environments, take the necessary
precautions (such as wearing gloves, clean room clothing, etc.) when handling
the stage as to avoid any contaminants. Maximum Bake-out temperature is 100°C.
MICRONIX USA supplies the stage with vacuum compatible connectors: 9-pin
female PEEK connectors for open loop, 15-Pin female PEEK connector for closed
loop with analog encoder, 25-pin female PEEK connector for closed loop with
digital encoder.
5.2.2
33B
Open loop Installation & Wiring Diagram
Connecting an open loop PPS-20 in a vacuum chamber requires the use of a feed
through connector at the vacuum chamber wall. The vacuum compatible PPS-20
will be supplied with wiring for a straight through feed through not a cross over
gender changer. MICRONIX USA supplies test connectors that simulate the
vacuum feed through to allow functionality test prior to installation in a vacuum
chamber. For details regarding the pin-out and feed through specifications see
the Appendix A.3.