5.2.3
Closed Loop/Encoder Installation & Wiring Diagram
Closed loop installation of the PPS-20 LMSM stage in vacuum environments
requires an intermediate feed through connector at the vacuum chamber wall
that can accommodate both the motor cable, and the encoder cable.
The vacuum compatible PPS-20 LMSM stage will be supplied with wiring for a
straight through feed through, not a cross over gender bender. MICRONIX USA
supplies test connectors that simulate the vacuum feed through to allow for
functionality testing prior to installation in a vacuum chamber. For details
regarding the pin-out and feed through specifications see the Appendix A.5.7,
A.6.5.
5.2.3.1
Analog Encoder Wiring Diagram
5.2.3.2
MII 6000 Digital Encoder Wiring Diagram
Stepper Motor Version
Stepper Motor Version
Temporary test
connectors are
standard. Requires
customer supplied
feed through
connector
Temporary test
connectors are
standard. Requires
customer supplied
feed through
connector