Conditions Table
Enter up to five stages of the following degas con-
ditions.
Soak Temperature
Enter the temperature at which the sample is to
soak while gas is flowing through the heating sta-
tion.
Range:
20 to 450
o
C
Default: 30
o
C
Temperature
Ramp Rate
Enter the rate at which the temperature will
change when advancing to the soak temperature.
Range:
5 to 20
o
C per minute
Default: 10
o
C per minute
Soak Time
Enter the amount of time the sample will soak.
Range:
0 to 10,000 minutes
Default: 10 minutes
Select a row (highlight the row), then press this
button to insert a row above the one selected.
Select a row (highlight the row), then press this
button to delete the row from the conditions table.
Press this button to clear the conditions table.
The rows do not have to be selected.
Press this button to save the information you
entered.
Press this button to close the Degas Conditions
dialog box.
Insert
Delete
Clear
Save
Close
SmartPrep
Options Menu
Aug 98
3-35
Содержание SmartPrep 065
Страница 1: ...SmartPrep 065 Operator s Manual V2 02 065 42801 01 March 2002...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 26: ......
Страница 28: ......
Страница 60: ......
Страница 61: ...CHAPTER 3 SMARTPREP MENUS File Menu SmartPrep Menu Reports Menu Options Menu...
Страница 62: ......
Страница 100: ......
Страница 101: ...CHAPTER 4 MAINTENANCE...
Страница 102: ......
Страница 105: ...CHAPTER 5 ORDERING INFORMATION...
Страница 106: ......
Страница 108: ......
Страница 109: ...APPENDIX A FORMS...
Страница 110: ......
Страница 112: ......
Страница 114: ......
Страница 115: ...APPENDIX B ERROR MESSAGES...
Страница 116: ......
Страница 127: ...INDEX...
Страница 128: ......