Verifying the Flow Rate
The flow rate for each port of the SmartPrep should be verified every 3 to 6
months to ensure a proper flow of gas.
1.
Insert the flow tube from port 1 into a beaker of water.
2.
Select the appropriate Unit [n] from the SmartPrep menu.
3.
Select Show Status from the Unit [n] menu; the following dialog is
displayed.
4.
Click
for Port 1. (Note that this push button changes to
.)
5.
Observe the liquid in the beaker.
•
If there is a steady stream of bubbles, the flow rate is appropriate.
•
If the bubbles are sparse and erratic, turn the flow control knob for port
1 to the right (clockwise).
•
If the liquid appears too bubbly, turn the flow control knob for port 1 to
the left (counterclockwise).
6.
Click
when the flow rate is as desired.
7.
Repeat steps 1 through 6 for each port.
Start gas flow
Stop gas flow
Stop gas flow
Maintenance
SmartPrep
4-2
Oct 01
Содержание SmartPrep 065
Страница 1: ...SmartPrep 065 Operator s Manual V2 02 065 42801 01 March 2002...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 26: ......
Страница 28: ......
Страница 60: ......
Страница 61: ...CHAPTER 3 SMARTPREP MENUS File Menu SmartPrep Menu Reports Menu Options Menu...
Страница 62: ......
Страница 100: ......
Страница 101: ...CHAPTER 4 MAINTENANCE...
Страница 102: ......
Страница 105: ...CHAPTER 5 ORDERING INFORMATION...
Страница 106: ......
Страница 108: ......
Страница 109: ...APPENDIX A FORMS...
Страница 110: ......
Страница 112: ......
Страница 114: ......
Страница 115: ...APPENDIX B ERROR MESSAGES...
Страница 116: ......
Страница 127: ...INDEX...
Страница 128: ......