Installing the AquaPrep
AquaPrep 055
2-2
Oct 09
4. Notify the Micromeritics Service Department of the defect and request shipping
instructions. The service department will assign a Returned Materials Authorization
(RMA) number. Write the RMA number on the outside of the shipping carton.
Installing the AquaPrep
Ideally, the AquaPrep should be placed next to the liquid sample handling module of the
DigiSizer (illustrated below). This allows for easy transfer of the container of deaerated water
from the AquaPrep to the sample handing module.
Selecting the Input Power
All instruments leave the factory set for 120 VAC and with the line fuse removed. The correct
setting of the universal power entrance must be checked and the appropriate fuse installed
before the AquaPrep can be operated. The AquaPrep is designed to operate with 100, 120,
220, or 240 VAC at 50 or 60 Hz. Voltage selection and fusing are made at the power connector,
which is located on the rear panel of the unit.
Do not connect the power cord until after you have verified the input power.
1. Make sure the power cord is not connected to the AquaPrep.
AquaPrep
Saturn DigiSizer
Liquid Sample Handler
The power cord should be disconnected from the AquaPrep before removing the
cover from the input power connector. Failure to disconnect the power cord could
result in electrical shock.
Содержание AquaPrep 055
Страница 1: ...AquaPrepTM 055 Operator s Manual 055 42801 01 October 2009...
Страница 4: ......
Страница 6: ......
Страница 20: ......
Страница 22: ......