EN
14
WatchBP Office Central
6)
Three consecutive measurements –
The device will
take three consecutive measurements with one minute
intervals between measurements.
A 60-second countdown before each measurement.
60
seconds
60
seconds
1
2
3
7)
The average measurement value – The average
blood pressure value will be displayed after the
measurements.
If one of the three measurements fail, a 4
th
measurement will be
conducted.
If the 3
rd
measurements is cancelled by pressing the Start/
Stop after the 2
nd
measurement, the averages are calculated
and saved to the memory as well. Press the <<M>> button to
display the average value. Press the <<M>> button again to
see the 2nd and 1st measurement again.
60
seconds
Taking less than three measurements
The measurement procedure can be stopped at anytime
by pressing the Start/Stop button in
«SCREEN» mode.
The device enters Standby mode and the remaining
measurements are cancelled. The average of the
measurements is saved and can be viewed by pressing the
<<M>> button.
Cancel remaining
measurements at anytime
during the measurement
sequence.
F
F