Acoustic:
The unit is turned ON and ready for the measure-
ment: 1 short beep.
Complete the measurement: 1 long beep (1 sec.) if
the reading is less than 37.5 °C / 99.5 °F, 10 short
«beep» sounds, if the reading is equal to or greater
than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi» sounds.
Memory:
30 readings recall in the memory mode with a
record of both time and date.
Backlight:
The display light will be GREEN for 1 second,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a reading
less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when
a measurement is completed with a reading equal
to or higher than 37.5 °C / 99.5 °F.
Operating
conditions:
Body mode: 15°C
- 40.0 °C / 59°F - 104.0 °F
Object mode: 5°C
- 40.0 °C / 41°F - 104.0 °F
15 - 95 % relative maximum humidity
Storage
conditions:
-25°C - +55 °C / -13°F - +131 °F
Automatic
Switch-off:
Battery:
Battery
lifetime:
15 - 95 %
%
relative maximum humidity
Approx.
after last measurement has been
taken.
2 x 1.5 V alkaline batteries; size AAA
approx. 1500 measurements
(using new batteries)
Dimensions:
1 x 4 x mm
Weight:
129g (with batteries),
g (w/o batteries)
IP Class:
IP22
Protected against solid foreign objects of 12.5mm
Ø and greater. Protected against vertically falling
This device complies with the requirements of the Medical
Device
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.
www.microlife-asiapacific.com
Expected
service life:
5 years or 12000 measurements
Reference to
standards:
water drops when enclosure tilted up to 15°.
ISO 80601-2-56,
IEC 60601-1, IEC 60601-1-2 (EMC),
IEC 60601-1-11