6

6 Read the recorded temperature from the LCD display.

7. For the next measurement, press the START/IO button 

4

 again to go back to the «ready for measuring» mode

9

, or wait 5 sec when «°C» or «°F» icon is flashing 

again and follow steps 4-5 above.

8. Press and hold the START/IO button 

4

 for 3 seconds to turn

off the device; otherwise the device will automatically switch off

after approx. 30seconds.

Measuring in object mode  

1.

Press the START/IO button 

4

. The display 

3

 is activated to 

show all segments for 1 second.

2.

Press the MODE button 

6

 to switch to object mode

.

3.

A flashing «

°C

»/«

°F

» icon, the blinking blue tracking light 

2

 and 

a beep indicate that the device is ready for measurement 

9

.

4.

Aim the thermometer at the center of the object you want to 

measure with a distance of no more than 5 cm. Press the 

START/IO button 

.

 After 

1

 seconds a long beep will verify 

the completion of measurement.

5.

Read the recorded temperature from the LCD display.

6.

For the next measurement, press the START/IO button 

4

 

again to go back to the «ready for measuring» mode 

9

, or 

wait 5 sec when «°C» or «°F» icon is flashing again and follow 

steps 4-5 above.

7.

Press and hold the START/IO button 

4

 for 3 seconds to turn 

off the device; otherwise the device will automatically switch off 

after approx. 30 seconds.

F

NOTE:

Patients and thermometer should stay in similar room

condition for at least 30 minutes.

Don't take a measurement while or immediately after nursing a

baby.

Don't use the thermometer in high humidity environments.

Patients should not drink, eat or exercise before/while taking

the measurement.

Don't move the measurement device from the measuring area

before hearing the termination beep.

10 short beeps and a red LCD backlight alert the patient that he/

she may have a temperature equal to or higher than 37.5 °C.

Always take the temperature in the same location, since

temperature readings may vary according to locations.

Doctors recommend rectal measurement for newborn infants 

within the first 6 months, as all other measuring methods might 

lead to ambiguous results. If using a non contact thermometer 

on those infants, we always recommend verifying the readings 

with a rectal measurement.

In the following situations it is recommended that three temper-

atures are taken with the highest one taken as the reading:

1. Children under three years of age with a compromised 

immune system and for whom the presence or absence of

fever is critical.

2. When the user is learning how to use the thermometer for 

the first time until he/she has familiarized himself/herself with

the device and obtains consistent readings.

3. If the measurement is surprisingly low.

Readings from different measuring sites should not be

compared as the normal body temperature varies by 

measuring site and time of day

, being highest in the evening

and lowest about one hour before waking up.

Normal body temperature ranges:

- Axillar: 34.7°C - 37.3 °C / 94.5°F - 99.1 °F

- Oral: 35.5°C - 37.5 °C / 95.9°F - 99.5 °F

- Rectal: 36.6°C - 38.0 °C / 97.9°F - 100.4 °F

- Microlife NC 300: 35.4°C - 37.4 °C / 95.7°F - 99.3 °F

9. Changing between Celsius and Fahrenheit

This thermometer can display temperature readings in either Fahr-

enheit or Celsius. To switch the display between °C and °F, 

press 

and hold

 the MODE button

6

 for 3 seconds; the beeper icon is 

shown on the display. Press the MODE button

6

 again; the current 

measurement scale («

°C

» or «

°F

» icon) will be shown on the 

display 

AN

. Change the measurement scale between °C and °F by 

pressing the M-button

5

. When the measurement scale has been 

chosen, press the START/IO button 

4

 to enter the  «ready for 

measuring» mode

9

; otherwise the device automatically switches to 

«ready for measuring» mode 

9

 after 10 seconds .

10.How to recall 30 readings in Memory Mode

This thermometer can recall the last 30 readings with a record of 

both time and date.

Recall mode 

AO

Press the M-button

5

 to enter recall mode

when the power is off. The memory icon «

M

» will display.

Содержание NC300

Страница 1: ...IB NC FR1MY1 EN 1122 Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com NC300 Non Contact Thermometer Microlife NC 300 1 2a 2b 3 4...

Страница 2: ...rolife NC 300 BP W100 12 AR BP W100 12 AR FA 9 AT AK AL AM AN AO AP AQ AR AS BT BK BL BO BP BM BN 2 1 3 4 6 8 5 7 NC 300 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Deal...

Страница 3: ...this thermometer Please read this document thoroughly before using the device and keep for future reference Type BF applied part Caution Recycling Batteries and electronic devices must be disposed of...

Страница 4: ...ction incorporates an advanced infrared sensor ensuring that each measurement is accurate and reliable Gentle and easy to use The ergonomic design enables simple and easy use of the thermometer This t...

Страница 5: ...indicated in the Technical Specifications section The residual risks and undesirable side effects that may be related to the use of this device can include but may not be limited to the following Devi...

Страница 6: ...If no further action is taken within 30 seconds the device will automatically turn off F Change current date and time Press and hold the MODE button 6 for approx 8 seconds until the year number start...

Страница 7: ...temperature readings may vary according to locations Doctors recommend rectal measurement for newborn infants within the first 6 months as all other measuring methods might lead to ambiguous results I...

Страница 8: ...C 32 2 F 211 8 F Resolution 0 1 C F Use an alcohol swab or cotton tissue moistened with alcohol 70 Isopropyl to wipe surface pollutions off the thermometer For cleaning the probe always start wiping t...

Страница 9: ...7 5 C 99 5 F Operating conditions Body mode 15 C 40 0 C 59 F 104 0 F Object mode 5 C 40 0 C 41 F 104 0 F 15 95 relative maximum humidity Storage conditions 25 C 55 C 13 F 131 F Automatic Switch off Ba...

Отзывы: