EN
29
WatchBP O3
"Er 13"
Cuff pressure
errors during
central blood
pressure
measurement
An adequate pressure
cannot be generated
in the cuff. A leak
may have occurred.
Check that the cuff is
correctly connected
and is not too loose.
Replace the batteries if
necessary. Repeat the
measurement.
"Er 15"
Abnormal
result of
central blood
pressure
reading
The measuring signals
are inaccurate and no
result can therefore
be displayed. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.
"Er F"
The device
has gone into
“single fault
condition”
Single fault condition
means that the
measurement is aborted
to protect the patient
from being harmed or
the device from being
damaged.
Re-position the cuff and
repeat the measurement.
Replace the batteries
if necessary. If the error
persists, contact microlife
or the local distributor.
"Er A"
Flash memory
error
Possible hardware fault.
Try again. If the error
persists, contact Microlife
or the local distributor.