WatchBP O3
Error messages
If an error occurs during measurement, the measurement
is interrupted and an error message «
Er
» is displayed.
• The device will take another
measurement automatically
when an error occurs.
• Please consult your doctor,
if this or any other problem
occurs repeatedly.
• If you think the results are
unusual, please read through
the information in this
instruction manual carefully.
Error
Description
Potential cause and
remedy
"Er 1"
Signal too
weak
The pulse signals on
the cuff are too weak.
Reposition the cuff and
repeat the measurement.
"Er 2"
Error signal
During the measurement,
error signals were
detected by the cuff,
caused for instance by
movement or muscle
tension. Repeat the
measurement, keeping
your arm still.