28
"Er 3"
No pressure in
the cuff
An adequate pressure
cannot be generated
in the cuff. A leak
may have occurred.
Replace the batteries if
necessary. Repeat the
measurement.
"Er 5"
Abnormal
result
The measuring signals
are inaccurate and no
result can therefore
be displayed. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.
"Er 11"
Signal too
weak during
central blood
pressure
measurement
The pulse signals on
the cuff are too weak.
Re-position the cuff and
repeat the measurement.
"Er 12"
Error signal
during central
blood pressure
measurement
During the measurement,
error signals were
detected by the cuff,
caused for instance by
movement or muscle
tension. Repeat the
measurement, keeping
your arm still.