Adjustment and Maintenance
4
Loosen the English 4-40 clamp screws on the Helical coupling between the
motor and the Ferrofluidic vacuum feedthrough by reaching through the
access hole in the motor mount. Remove the three ¼-28 English screws which
attach the motor mount to the Conflat vacuum flange on the service ring.
Remove the motor mount assembly (including motor, Helical coupling, travel
lock plate, and encoder) from the Conflat vacuum flange, leaving the Conflat
flange and Ferrofluidic vacuum feedthrough attached to the service ring.
Inside the service ring, loosen the socket head clamp screw on the Helical
shaft coupler where it attaches to the lower drive pulley shaft. Remove the
three remaining Conflat mounting screws. Remove the Conflat with
Ferrofluidic feedthrough attached.
Remove the Helical coupling from the feedthrough, and unscrew the
feedthrough from the Conflat flange.
Lubricate the O-ring on the new Ferrofluidic vacuum feedthrough with a
light coat of Apiezon L grease. Use pliers with padded jaws (e.g. blue pad) to
gently tighten the feedthrough to the Conflat flange.
Reverse the procedure for reassembly.
4.1.1.6.Replacing the V-Plate
The V-plate contains three tungsten V's which support the test mass. Since
removal and insertion of the tungsten V's in the V-plate requires special tools,
this cannot be done in the field. However, an entire new V-plate assembly
can be installed. The dropping chamber must be vented and opened for this
procedure. If possible, use dry nitrogen to vent the chamber.
4-4
Remove the bottom drag-free cover
from the cart by removing the three M3
screws. Detach the LED bracket from the side of the cart and pull it out of the
way, being careful not to detach or damage the wires connected to it.
Remove the two M3 screws and the threaded post attaching the top drag-free
Содержание FG5
Страница 19: ...Design Components and Function 2 Figure 2 8 Top view of interferometer optics and beam path 2 11 ...
Страница 20: ...Design Components and Function 2 Figure 2 9 Side view of interferometer optics and beam path 2 12 ...
Страница 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Страница 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Страница 32: ......
Страница 42: ...How to Set Up and Run the FG5 3 3 10 ...
Страница 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Страница 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Страница 87: ...Adjustment and Maintenance 4 4 29 ...
Страница 91: ...Adjustment and Maintenance 4 4 33 ...
Страница 104: ...Troubleshooting 5 5 2 ...
Страница 117: ...Troubleshooting 5 5 15 ...
Страница 131: ...Checklists and Logs Appendix D 9 3 ...
Страница 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Страница 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Страница 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...