Name
Description
Documentation Text
Measurement Control
GraycodeThreshold
Minimum allowed pattern
contrast in the graycode
images (for 3D reconstruc-
tion);
Only pixels whose gray value difference between the bright and dark
images is equal to or greater than the limit value set are used for the
calculation.
AmplitudeThreshold
Minimum allowed pat-
tern contrast in the sinus
images
(for 3D reconstruction);
Only pixels whose amplitude value is greater than this threshold value
are used to calculate the result. This allows you to exclude unwant-
ed pixels with a low degree of reflection (e.g., outside the measured
object or near the edges).
MaximumPhaseGradientEnable Controls if threshold to
sort out outliers is active.
Enables filtering of high gradient data in Z.
MaximumPhaseGradient
Threshold to sort out outli-
ers in 3d point cloud.
Threshold for filtering with high gradient in Z.
MinimumIntensityThreshold
Threshold to sort out
underdriven pixels.
Threshold for filtering of underxposed pixels.
MaximumIntensityThreshold
Threshold to sort out over-
driven pixels.
Threshold for filtering of overexposed pixels.
PatternWidth
Sets the pattern period for
sine pattern in pixel.
Strip width of the projected sine pattern.
PatternCount
Sets the number of
patterns within one sine
pattern sequence.
The number of sine patterns and images to be recorded which are
used for one measurement.
PatternType
Selects the type of pattern
projection.
Provides pre-defined options for setting the number of sine patterns
which are used for one measurement:
- HighSpeed: 4
- Balanced: 6
- HighPrecision: 12
PatternType
Selects the type of pattern
projection.
Provides pre-defined options for setting the number of sine patterns
and the images to be recorded that are used for one measurement:
- HighSpeed: 4
- Standard: 12
- HighPrecision: 24
- Custom: Enables the user-specific PatternWidth and PatternCount
registers
PerformanceMode
Boosts data processing
speed while increasing
internal temperature of
sensor.
Significantly increases the data processing speed at the expense of
the sensor’s internal temperature.
- On: The measurement speed will increase depending on the sensor
settings by up to 85 %.
- Off: The sensor’s internal temperature can decrease by up to approx.
5 °C during continuous operation.
The sensor can thus be operated in correspondingly higher ambient
temperatures.