Page 67
Operation
English
optoCONTROL 1200/1201
6.8
Analog Output
Receiver fully
exposed
10 V
V
max
0 %
Measuring
range
max.
Signal
Signal
Signal
0 %
Measuring
range
max.
0 %
Measuring
range
max.
Receiver completely
blocked
Target
Light source
Receiver
The system outputs
V
max
if no
object blocks any light and
the receiver is fully exposed.
The voltage
V
max
can be set via
the gain potentiometer and
typically should read 10 V.
As the light for the receiver
decreases the signal drops
linear according to light
quantity.
If the light curtain is fully cov-
ered by a target the receiver
outputs a signal of 0 V.
i
Take care on target orien-
tation to avoid measure-
ment errors!
Fig. 14 Analog signal behavior with different target positions
Output gain
V
OUT
10.0 V = 100 % measuring range (no target)
MR
= Measuring range
x
[mm] =
MR
[mm] * (10 V -
V
OUT
)
10 V
Fig. 15 Calculation of a value
Example:
V
OUT
= 4.6 V
Measuring range = 10 mm
Value
= 5.4 mm