Page 46
Operation
induSENSOR MSC 7210
1. Set the frequency with J4 ein, see Chap.
5.2.1
.
2. Set the phase with J2, see Chap.
5.2.1
.
3. Set the supply voltage for the sensor with J1, see Chap.
5.2.2
.
Plunger resp. the sensor can be adjusted mechanical
during the settings (example A, see Chap.
5.3
.)
4a:
Move switch S1 at J3 to OFF and S2 at J3 to ON,
see
Fig. 12
.
Initially the plunger is situated outside of the sensor.
5a:
Set the desired output signal with R1 for the
sensor in middle-position, see
Fig. 11
.
6a:
Move plunger in middle position and move until
the desired output signal is reached (see 5a).
Magnetic core is in middle position (electrical and me-
chanical zero point of the sensor).
7a:
Displace plunger respectively the sensor into
desired end position, use R3 to set the amplification.
8a:
Displace the plunger resp. the sensor to the
position as described in 6a. Readjust output signal
with R1 to the value defined in 5a.
9a:
Move plunger again into desired end position
and readjust the amplification with R3.
End of setting.
Plunger resp. the sensor is fixed to the target during
the settings (example B, see Chap.
5.3
).
4b:
Move switch S1 and S2 at J3 to ON, see
Fig. 12
.
5b:
Set the desired output signal with R1 for the
start position, see
Fig. 11
.
6b:
Move switch S1 and S2 at J3 to OFF.
7b:
Displace the plunger resp. the sensor to the
start position. Use R2 to set the output to the
same value as in point 5b, see
Fig. 11
.
8b:
Displace the plunger resp. the sensor to
the end position and use R3 to set the desired
signal for the end position.
End of setting.