Page 72
Default Setting
optoNCDT 1402
Correct
laser on
state
teach in
opto
NCDT
Incorrect
(shadow)
nach DIN EN 60825-1: 2007
P
x
1 mW;
O
= 670 nm
LASERSTRAHLUNG
NICHT IN DEN STRAHL BLICKEN
LASER KLASSE 2
In case of bore holes, blind holes, and edges in
the surface of moving targets the sensor must
be arranged in such a way that the edges do not
obscure the laser spot (Fig. 24).
Sensor arrangement for holes and ridges
Fig. 24:
Cleaning the Protective Glasses
9.4
Clean the protective glass with high-percentage alcohol (propanol or optics cleaner) and a lint-free lens
cleaning paper or an optics cleaning cloth (from the optics or photo shop).
The surface is damaged if a dry cloth is used for cleaning.
Default Setting
10.
Data protocol ILD1402
-
Current output with error value (3.75 mA)
-
Measuring rate : 1.5 kHz
-
Interface: 115.2 kBaud, binary format (no ASCII)
-
Moving average avg =1 (no averaging)
-
Teach value 1: 0.0
-
Teach value 2: 16368.0
-
External input for scaling
-
Continuous measurement output
-
Output time: 500 ms,
-
Key lock after 5 min power on
-
Settings saved into FLASH
-
Measuring range:
-
100 % FSO: I = 20 mA , digital 16207
0 % FSO: I = 4 mA, digital 161
Maximum output (101 % FSO):
-
20.16 mA / digital 16367
Minimum output (-1 % FSO):
-
3.84 mA / digital 0
Set sensor on default settings:
Switch off the senors power supply.
Keep the key “Select“ pressed.
Switch on the sensors power supply.
Wait about 5 sec.