Page 53
Serial Interface RS422
optoNCDT 1402
Reply:
31
24 23
16 15
8 7
0
hex
„l“
„L“
„D“
„1“
0x494C4431
0xA0
0x90
0x00
0x02
0xA0900002
0x20
0x20
0x0D
0x0A
0x20200D0A
Characteristics for Digital or Analog Data Output
8.3.9
Name:
SET_OUTPUTMODE
Description:
Sets the output characteristics.
X = 0 -->
-
continuously each measurement, depending on baud rate and the measuring frequency;
delay = (Bit quantity / Baud rate) * measuring frequency [Hz], if delay < 0, delay = delay +1)
delay = number of cycles with no serial output
X = 1 --> time-based, see Chap. 8.3.10.
-
X = 2 --> trigger controlled, see Chap. 8.3.14
-
Format: 31
24 23
16 15
8 7
0
hex
„+“
„+“
„+“
0x0d („CR“)
0x2B2B2B0D
„l“
„L“
„D“
„1“
0x494C4431
0x20
0xF4
0x00
0x03
0x20F40003
0x00
0x00
0x00
0x0X
0x0000000X
Reply:
31
24 23
16 15
8 7
0
hex
„l“
„L“
„D“
„1“
0x494C4431
0xA0
0xF4
0x00
0x02
0xA0F40002
0x20
0x20
0x0D
0x0A
0x20200D0A
Default setting:
continuously