thermoMETER CTVideo
31
All alarms (alarm 1, alarm 2, output channel 1 and 2 if used as
alarm output) have a fixed
hysterese of 2 K
.
Both of these alarms will have
effect on the LCD color:
BLUE:
alarm 1 active
RED:
alarm 2 active
GREEN:
no alarm active
Alarms
The CTVideo has the following Alarm
features:
Output channel 1
To activate the alarm function the output channel has to be switched into digital mode. For this purpose the
software is required.
Visual Alarms
These alarms will cause a change of the color of the LCD display and will also change the status of the
optional relays interface. In addition the Alarm 2 can be used as open collector output at pin
AL2
on the
mainboard
[24 V/ 50 mA]
.
From factory side the alarms are defined as follows:
Alarm 1
Norm. closed/ Low-Alarm
Alarm 2
Norm. open/ High-Alarm
Extended setup like definition as low or high alarm
[via change of normally open/ closed]
, selection of the
signal source
[TObj, THead, TBox]
can be done with the software.