thermoMETER CTVideo
25
Designation
+8
… 36 VDC
Power supply
GND
Ground (0 V) of power supply
GND
Ground (0 V) of internal in- and outputs
AL2
Alarm 2 (Open collector output)
OUT-TC
Analog output thermocouple (J or K)
OUT-mV/mA
Analog output object temperature (mV or mA)
F1-F3
Functional inputs
GND
Ground (0 V)
3V SW
BLACK
/ Power supply Laser (+)
GND
GREY
/ Ground Laser (
–)
ORANGE
Temperature probe sensor (NTC)
BLUE
Sensor ground
PURPLE
Sensor power
YELLOW
Detector signal
Opened controller
with terminal connections
Power supply
Please use a separate, stabilized power supply unit with an output voltage of
5
– 28 VDC
which can supply
160 mA.
The ripple should be max.
200 mV
.
Please use shielded cables only for all power and data lines.
The sensor shield has to be grounded.
CAUTION: Please do never connect a supply voltage to the analog outputs as this will destroy the output!
The CTVideo is not a 2-wire sensor !