thermoMETER CTVideo
28
Sensor head Calibration Code
Every sensor has a specific calibration code, which is printed on the sensor. For a correct temperature
measurement and functionality of the sensor this calibration code must be stored into the controller.
The calibration code consists of five blocks with 4 characters each.
Example:
EKJ0
– 0OUD – 0A1B – A17U – 93OZ
block1 block2 block3 block4 block5
For entering the code please press the
Up
and
Down
key (keep pressed) and then the
Mode
key.
The display shows
HCODE
and then the 4 signs of the first block. With
Up
and
Down
each sign can be
changed,
Mode
switches to the next sign or next block.
After you have modified the sensor code a reset is necessary to activate the change.
[► Operating]
You will find the calibration code on a label fixed on
the sensor. Please do not remove this label or make
sure the code is noted anywhere.